1.Package table heat slowly.
The fixing screw of the heating plate on the packaging table is loose or the heating plate is aging.
2.The measured temperature of the packaging table is different from that shown by the computer.
Use a flat-head screwdriver to adjust the VR1 adjustable resistance on the circuit board
(turn up the temperature clockwise and turn down the temperature counterclockwise).
3.Package table hit taper table.
a.Check whether the width of the package table in the system setting is 2mm larger than the actual width of the package table.
b.Whether the base of the package table is in the middle of the tapering table.
4.Flowmeter leakage.
Turn off the output button of the flowmeter, remove the gas pipe on the side of the heater, fill a cup of water with a beaker, and put the output pipe of the flowmeter into the water. If the flowmeter leaks, bubbles will appear, then adjust the zero port on the inlet pipe of the flowmeter counterclockwise with a small flat-head screwdriver. When adjusting, pay attention to slowly adjust, until the output pipe is just not leaking, the computer displays the output flow rate in 0-2SCCM (First of all, it is subject to no leakage,and if it is adjusted more, the output flow will be smaller than the set flow).
5.“Illegal Characters Exist” is displayed during parameter is selected into use.
The parameter Settings cannot contain Spaces, periods, and commas.